Automated Blood Culture Detection System
Instrument Features & Specifications
Test Principles
Colorimetry & CO2 sensor
Sample Type
Blood and various sterile body fluid specimens
Capacity
64 bottles
Detection Method
Continuous swing vibration
An independent detector in each position, providing real-time status
Detect the signal every 10 minutes, 24 hours non-stop monitoring
90% positive samples can be obtained within 24 hours, 98% positive samples can be obtained within 48 hours
Interpretation Method
Sustained acceleration, Rate, and Initial threshold
Sample Loading
Able to accept samples in random positions and at any time in each rack
Able to resume the procedure if a sample is picked up and reinserted within at least 2 hours
Able to process samples with a late arrival up to 48 hours
Working Environment
Environmental temperature: 5℃–37℃
Relative humidity: 10%–90%
Operation System & Software
Internal computer with process control, data storage and analysis, patient management, and exam reporting capability
Real-time identification and warning, with audio, visual, and software alarm, of positive samples
Atmospheric Pressure
76kPa–106kPa
Power Supply
AC 110-240V, 50/60 HZ
Dimension & Weigh
575mm * 560mm * 950 mm; 80kg
Blood Culture Bottle Features & Specifications
Bottle Type
Adult aerobic blood culture bottle
Adult anaerobic blood culture bottle
Children’s blood culture bottle
Package
40 bottles/box Box size:282*177*127mm G.W.:2.6 kg
6 boxes/carton Carton size:500*300*275mm G.W.:16.1kg
Bottle Design
Made of Multilayer polymeric fibers
Lightweight and unbreakable, avoid biohazard
A special formula of media to improve the cultivation of fastidious bacteria
Resin microsphere guarantees antibiotics absorbing, prevents antibiotics interference; It lyses WBC and release bacteria inside to improve detection rate; Microsphere also provides enough surface for bacteria growing;
Colorless resin design completely avoids interference of gram stain against activated carbon
Resin Microspheres has a great advantage over active carbon, which is fully reflected in smear staining to positive blood culture specimen. The left one shows the specimen smear of an active carbon bottle, so we can see the background with few clouds of black that must influence on observation on smear and seeking bacteria. The right one shows the specimen smear of the resin culture bottle, which is almost transparent, so the visual field of it is very clear and the bacteria can be found easily.