LCD Screen Single Channel Syringe Pump
Special Features
1. 3.5 inch LCD screen for clear vision within 5 meters
2. Especially designed extension line grip
3. One-button night mode design for patient’s more comfortable rest at night
4. Compatible with different brands of 5,10,20/30,50/60 ml syringes
5. Free stacking of single syringe pumps for easy space stations in ICU, NICU, OT, etc
| Model No. Parameters | LTSI24 Single Channel Syringe Pump | |
| Channel | Single | |
| Stackable | YES | |
| Syringe Size | 5,10,20,30,50/60ml | |
| Infusion Modes | Rate Mode, Rate-Time, Rate-VTBI, Time-VTBI, Body Weight | |
| Safety | Double CPU to make a sure higher standard infusion | |
| Drug Library | 20 drug list with drug code display | |
| Accuracy | ±2% | |
| Time Preset | 00: 01~99: 59 (Hour: Min) | |
| Volume Range | 0~9999.9ml | |
| Flow Rate Step by | 0.1 ml/h when Rate<100ml/h, 1ml/h when Rate ≥100ml/h | |
| Flow Rate Range | 5ml Syringe 0.1ml/h-150ml/h | |
| 10ml Syringe 0.1ml/h-300ml/h | ||
| 20ml Syringe 0.1ml/h-600ml/h | ||
| 30ml Syringe 0.1ml/h-900ml/h | ||
| 50/60ml Syringe 0.1ml/h-1200ml/h | ||
| Purge/Bolus Rate/Top Flow Rate | 5ml Syringe 150ml/h | |
| 10ml Syringe 300ml/h | ||
| 20ml Syringe 600ml/h | ||
| 30ml Syringe 900ml/h | ||
| 50/60ml Syringe 1200ml/h | ||
| Audible & Visible Alarms | Auto Self-Test, Syringe Dislocation, Occlusion, Near-End, Syringe Empty, VTBI Completion, Low Battery, Battery Exhausted, Motor Malfunction, Incorrect Syringe Specification, Circuit Malfunction, Master CPU Malfunction, Monitoring CPU Malfunction, Parameters Over Limitation, A.C Disconnection, A.C connection | |
| KVO | 0.1-5.0ml/h Adjustable | |
| Occlusion Pressure | High | 40 KPa±20KPa | 
| Middle | 60 KPa±20KPa | |
| Low | 100KPa±20KPa | |
| Max-Infusion Pressure | 120KPa | |
| Battery | ≥8 hours | |
| Power Consumption | 30VA | |
| Power Supply | AC100-240V,50Hz/60Hz | |
| Battery | Lithium battery, 11.1/2000mAh | |
| Classification | Class II, type CF, IPX4 | |
| Dimension | 26×21.5×275px | |
| Weight | 2kg | |
| Optional Function | Ambulance DC 12V | |